General Specifications
- Detection Modes:
- UV-Vis Absorbance
- Fluorescence Intensity
- Luminescence
- Time-Resolved Fluorescence (Secondary Mode)
- Alpha Detection
- Microplate Types: Supports 6- to 384-well plates
- Read Methods: End point, kinetic, spectral scanning, and well area scanning
- Other Labware Supported: PCR plates, Petri and cell culture dishes, Take3 Micro-Volume Plates
- Temperature Control:
- 4-Zone™ Incubation up to 50°C with Condensation Control™
- Accuracy: ±0.2°C at 37°C
- Shaking Options: Linear, orbital
- Software: Gen5™ Microplate Reader and Imager Software
- Automation Compatibility: BioStack and 3rd party automation compatible
Absorbance Mode
- Light Source: Xenon flash
- Detector: Photodiode
- Wavelength Selection: Monochromator
- Wavelength Range: 200 – 999 nm (1 nm increments)
- Monochromator Bandwidth: 2.4 nm
- Dynamic Range: 0 – 4.0 OD
- Resolution: 0.0001 OD
- Pathlength Correction: Yes
- Wavelength Accuracy: ±2 nm
- Wavelength Repeatability: ±0.2 nm
- OD Accuracy: <1% at 2.0 OD, <3% at 3.0 OD
- OD Linearity: <1% from 0 to 3.0 OD
- OD Repeatability: <0.5% at 2.0 OD
- Stray Light: 0.03% at 230 nm
- Reading Speed (Kinetic):
- 96 wells: 14 seconds
- 384 wells: 26 seconds
Fluorescence Intensity Mode
- Light Source: Tungsten halogen (Xenon flash optional)
- Detector: PMT
- Wavelength Selection: Filters
- Wavelength Range: 300 – 700 nm (option for 200 – 850 nm)
- Dynamic Range: >6 decades
- Sensitivity:
- Top and Bottom: Fluorescein 5 pM (1 fmol/well, 96-well plate)
- Reading Speed (Kinetic):
- 96 wells: 31 seconds
- 384 wells: 80 seconds
Luminescence Mode
- Wavelength Range: 300 – 700 nm
- Dynamic Range: >6 decades
- Sensitivity:
- 10 amol ATP (Flash) – Lum and Abs/Lum configurations
- 30 amol ATP (Flash) – Multi-mode configurations
Time-Resolved Fluorescence Mode
- Light Source: Xenon flash
- Detector: PMT
- Wavelength Selection: Monochromator
Alpha Detection Mode
- Light Source: Tungsten halogen
- Detector: PMT
- Wavelength Selection: Filters
- Sensitivity: 300 amol of biotinylated-LCK-P peptide (25 µL/well in 384-well plate)
- Read Speed: 2 minutes (96-well plate)